Abstract
This paper reports on a radio frequency micro-strip T-resonator that is integrated to a parallel droplet microfluidic system. The T-resonator works as a feedback system to monitor uniform droplet production and to detect, in real-time, any malfunctions due to channel fouling or clogging. Emulsions at different W/O flow-rate ratios are generated in a microfluidic device containing 8 parallelized generators. These emulsions are then guided towards the RF sensor, which is then read using a Network Analyzer to obtain the frequency response of the system. The proposed T-resonator shows frequency shifts of 45MHz for only 5% change in the emulsion's water in oil content. These shifts can then be used as a feedback system to trigger alarms and notify production and quality control engineers about problems in the droplet generation process.