Abstract
Aim: The present work provides a fast, simple, accurate, and inexpensive analytical method for the determination of Linagliptin (anti-diabetic drug).
Methods: The analysis was performed using a square wave adsorptive anodic stripping voltammetric technique (SWAASV) and glassy carbon electrode (GCE) as a working electrode. The experimental and instrumental parameters were studied and discussed to ensure the validity of the method.
Results: The method has a very good linearity (R-2 = 0.9984), wide concentration range (0.189 - 2.268 mu g mL(-1)), low detection limit of 0.052 mu g mL(-1) and low quantitation limit of 0.172 mu g mL(-1).