Abstract
A simple and rapid reversed-phase high-performance liquid chromatographic method for monitoring the reactions involved in two different processes for the production of 4-methoxyphenylacetic acid (PMPA) was developed. Impurity profiles of PMPA were used for fingerprinting of the two different synthetic processes by HPLC. Impurities were separated and determined on a Hypersil C
18 column with acetonitrile–0.1
M potassium dihydrogen orthophosphate–triethylamine (40:59.95:0.05, v/v) (pH 3.0) as the mobile phase and detection at 280 nm at ambient temperature. The method was substantiated with respect to accuracy, precision, linearity, robustness, limit of detection and quantification. The method was found to be suitable not only for monitoring the reactions but also for quality assurance of PMPA as it could detect impurities at the level of 4·10
−9 g.