Abstract
The objectives of this study were to estimate patient dose in some common diagnostic X-ray examinations. Radiation doses were estimated for 307 patients in six public hospitals comprising 7 X-ray units in Wad-madani, Sudan. Entrance surface air kerma (ESAK) was estimated in a three step protocol: First, X-ray unit output Y(d) was measured at a distance, d for different peak tube voltages and tube loadings (mAs). Next, incident air kerma (K-i) was calculated from Y(d) using inverse square law combined with patient exposure factors. ESAK was calculated from K-i using backscatter factor, B. Mean ESAK values are comparable to those reported in other countries and are below reference dose levels. The estimated mean ESAK values are: 0.3, 2.2, 2.2, 2.9, 2.8, 3.1, and 7.5 mGy for chest PA, Skull AP/PA, Skull LAT, Abdomen, Pelvis AP, Lumbar Spine AP and Lumbar Spine LAT examinations, respectively. The results are used for dose optimization, and to propose local diagnostic reference levels.