Abstract
We report an evanescent wave based fibre optic absorbance meter that enables the colorimetric detection of waterborne cations with water insoluble chromoionophoric sensitisers. This establishes an alternative to the PVC membrane based transducers that are conventionally used for this purpose. Here, a water insoluble sensitiser is coated as a thin film on an unclad section of a multimode optical fibre to overlap with the evanescent field of a light beam propagating along the fibre core. The colorimetric response of the sensitiser when in contact with waterborne cation leads to increased absorption of virtual photons associated with the evanescent field. The resulting intensity loss of the propagating beam is detected by a bespoke self-referenced evanescent wave absorbance meter with beam intensity modulation and Lock-in amplification. We validate our transducer with the well characterised water insoluble sensitizer, 1-(2-pyridylazo)-2-naphthol (PAN), for the detection of aqueous Zn2+ cations. We find a limit of detection (LoD) of 54 nM Zn2+, 28 times lower compared to a PVC membrane based sensor using same sensitiser for same cation (Albero et al., Journal of Pharmaceutical and Biomedical Analysis 29 (2002), 779). Our evanescent wave absorbance meter can easily be adapted to other colorimetric sensitisers, including chromoionophoric complex forming macrocycles. (C) 2016 Elsevier B.V. All rights reserved.