Abstract
•The first HPLC-UV method for the determination of erdafitinib.•Application on pharmacokinetic studies in mice.•Optimization of SPE method for extracting erdafitinib from plasma with high recovery.•Full method validation according to FDA and EMA bioanalytical method validation guidelines.
Erdafitinib is a recently approved fibroblast growth factor receptor (FGFR) inhibitor. It is the first treatment targeting susceptible FGFR genetic alterations for patients with metastatic bladder cancer. A simple validated HPLC-UV method was developed for the determination of erdafitinib in mouse plasma. Erdafitinib and internal standard (rivaroxaban) were efficiently separated on Eclipse plus C18 column (4.6 × 100 mm, 3.5 µm). The mobile phase consisted of acetonitrile and 0.01 M ammonium acetate aqueous solution, adjusted to pH 4.4 with acetic acid (26:74, v/v) and it was eluted isocratically at a flow rate of 1.2 mL/min. The UV detection was at 292 nm and the total run time for each sample was 11 min. The method linearity was validated over the range of 0.05–2.00 µg/mL (r2 ≥ 0.9992) and the lower limit of quantification (LLOQ) was 0.05 µg/mL. The within-run and between-run accuracies were 98.56 and 99.24%, respectively while the CV of the method precision did not exceed 6.52%. Plasma samples were extracted using a solid phase extraction procedure and the extraction recoveries were 97.90 ± 4.58%. The method was optimized for the sensitive determination of the studied drug in mouse plasma and was successfully applied to its pharmacokinetic studies.