Abstract
The Hybrid Compact Gamma Camera (HCGC) is a small field of view (SFOV) portable hybrid gamma-optical camera intended for small organ imaging at the patient bedside. In this study, a thyroid phantom was used to determine the suitability of the HCGC for clinical thyroid imaging through comparison with large field of view (LFOV) system performance.
A direct comparison with LFOV contrast performance showed that the lower sensitivity of the HCGC had a detrimental effect on image quality. Despite this, the contrast of HCGC images exceeded those of the LFOV cameras for some image features particularly when a high-resolution pinhole collimator was used.
A clinical simulation showed that thyroid morphology was visible in a 5 min integrated image acquisition with an expected dependency on the activity within the thyroid. The first clinical use of the HCGC for imaging thyroid uptake of I-123 is also presented.
Measurements indicate that the HCGC has promising utility in thyroid imaging, particularly as its small size allows it to be brought into closer proximity with a patient. Future development of the energy response of the HCGC is expected to further improve image detectability. (C) 2015 Associazione Italiana di Fisica Medica. Published by Elsevier Ltd.