Abstract
Present interest concerns development of a system to measure photoelectron-enhanced dose close to a tissue interface using analogue gold-coated doped silica-fibre thermoluminescence detectors and an X-ray set operating at 250 kVp. Study is made of the dose enhancement factor for various thicknesses of gold; measurements at a total gold thickness of 160 nm (accounting for incident and exiting photons) produces a mean measured dose enhancement factor of 1.33 +/- 0.01 To verify results, simulations of the experimental setup have been performed. (C) 2013 Elsevier Ltd. All rights reserved.