Abstract
Background: Proper sealing of screw-access channels against microbial microleakage is advisable for the long-term success of screw-retained implant prosthesis. Objective: This study aimed to compare the bacterial adhesion and microleakage with three restorative materials, namely, composite resin, acrylic resin and bis-acryl, that are used to cover the access channels of screw-retained implant prostheses, using polytetrafluoroethylene tape as a spacer material. Materials and Methods: In this in vitro study, 18 titanium straight abutments (Hex-lock (R) Zimmer) were torqued into implant analogs, which were then subdivided into three groups. The samples of each group were filled with polytetrafluoroethylene tape and sealed with the three restorative materials (Group A: composite resin; Group B: acrylic resin; Group C: bis-acryl). Measurements of surface bacterial adhesion and internal microleakage were then recorded. The results were statistically analyzed using Kruskal-Wallis and Chi-square tests. Results: No significant difference was found between the investigated materials in terms of their sealing effectiveness against microbial microleakage (P = 0.06). Regarding bacterial adhesion, composite resin showed the highest number of surface adhesion, but there was no significant difference between the three materials (P = 0.081). Conclusion: The results of this study suggest that composite resin, acrylic resin and bis-acryl materials could be used alternatively in sealing the implant access channel owing to no significant differences in terms of microleakage and bacterial adhesion.