Abstract
Current sizing strategies suggest valve selection based on annulus diameter despite supra-annular placement of biological prostheses potentially allowing placement of a larger size. We assessed the frequency of selecting a larger prosthesis if prosthesis size was selected using a replica (upsizing) and evaluated its impact on haemodynamics.
We analysed all discharge echocardiograms between June 2012 and June 2014, where a replica sizer was used for isolated aortic valve replacement (Epic Supra: 266 patients, Trifecta: 49 patients).
Upsizing was possible in 71% of the Epic Supra valves (by 1 size: 168, by 2 sizes: 20) and in 59% of the Trifectas (by 1 size: 26, by 2 sizes: 3). Patients for whom upsizing was possible had the lowest pressure gradients within their annulus size groups. The difference was significant in annulus diameters of 21-22 or 25-26 mm (Epic Supra) and 23-24 mm (Trifecta). Trifecta gradients were the lowest. However, the ability to upsize the Epic Supra by 2 sizes eliminated the differences between Epic Supra and Trifecta. Upsizing did not cause intraoperative complications.
Using replica sizers for aortic prosthesis size selection allows the implantation of bigger prostheses than recommended in most cases and reduces postoperative gradients, specifically for Epic Supra.