Retrofitting an environmental monitor with a silicon photomultiplier sensor
Journal article   Open access  Peer reviewed

Retrofitting an environmental monitor with a silicon photomultiplier sensor

Mohammed Alharbi, Peter H Burgess, Matt Taggart, Paul J Sellin and M. K. M. Alharbi
Journal of radiological protection, Vol.40(3), pp.N31-N38
01/09/2020
PMID: 32408280

Abstract

environmental monitor silicon photomultiplier scintillator
We report on the retrofitting of a standard DP2 environmental radiation monitor replacing the photomultiplier tube with a silicon photomultiplier (SiPM). The use of a SiPM has several advantages for a hand-held radiation monitor, including convenient low voltage operation and physical robustness. The SiPM is used to replace the existing photomultiplier tube, and we report the detection efficiency and alpha/beta discrimination performance of the modified probe compared to an unmodified version.
url
https://doi.org/10.1088/1361-6498/ab9312View
Published (Version of record) Open

Metrics

1 Record Views

Details