Abstract
An on-line high-performance liquid chromatography (HPLC) method for the rapid and selective detection of boronic acids in complex mixtures was developed. After optimization experiments at an HPLC flow rate of 0.40 mL/min, the HPLC-separated analytes were mixed post-column with a solution of 75 μM alizarin and 0.1% triethylamine in acetonitrile, which was delivered at a flow rate of 0.60 mL/min. The reaction between alizarin and boronic acids occurred in a reaction coil of dimensions of 3.5 m × 0.25 mm at a temperature of 50 °C, resulting in fluorescent complexes that were detected as positive peaks by a fluorescence detector (λexc 469 nm and λe m 610 nm). The method enabled the selective detection of various boronic acids and derivatives, with a limit of detection of phenylboronic acid of 1.2 ng or 1 μM. It could successfully monitor the progress of two organic reactions involving boronic acid-containing compounds, and provided useful insights into the course of the reactions.