Abstract
A sensor for hydrogen peroxide is described that is based on an indium tin oxide electrode modified with Fe3O4 magnetic nanoparticles which act as a mimic for the enzyme peroxidase and greatly improve the analytical performance of the sensor. The amperometric current is linearly related to the concentration of H2O2 in the range from 0.2 mM to 2 mM, the regression equation is y = -0.5-1.82x, the correlation coefficient is 0.998 (n = 3), and the detection limit is 0.01 mM (S/N = 3). The sensor exhibits favorable selectivity and excellent stability.