Sign in
Study on fine-mesh PMTs for detection of aerogel Cherenkov light
Journal article   Peer reviewed

Study on fine-mesh PMTs for detection of aerogel Cherenkov light

T. Iijima, M. Amami, I. Adachi, R. Enomoto, R. Itoh, S. Kobayashi, T.H. Liu, D.R. Marlow, A. Murakami, S. Ogawa, …
Nuclear instruments & methods in physics research. Section A, Accelerators, spectrometers, detectors and associated equipment, Vol.387(1), pp.64-68
01/03/1997

Abstract

Most recent fine-mesh photomultipliers (FM-PMTs), with 19 dynode stages, have been tested for application to a threshold aerogel Cherenkov counter. Two properties, the gain and the pulse height resolution, have been particularly studied in a magnetic field of up to 1.5 T. The obtained results show that a FM-PMT is a suitable device for detection of aerogel Cherenkov light.

Metrics

1 Record Views

Details