Abstract
Objectives: To assess the accuracy of resting energy expenditure (REE) measurement in a sample of overweight and obese Saudi males, using the BodyGem device (BG) with whole room calorimetry (WRC) as a reference, and to evaluate the accuracy of predictive equations.
Methods: Thirty-eight subjects (mean +/- SD, age 26.8 +/- 3.7 years, body mass index 31.0 +/- 4.8) were recruited during the period from 5 February 2007 to 28 March 2008. Resting energy expenditure was measured using a WRC and BG device, and also calculated using 7 prediction equations. Mean differences, bias, percent of bias (%bias), accurate estimation, underestimation and overestimation were calculated.
Results: Repeated measures with the BG were not significantly different (accurate prediction: 81.6%; %bias 1.1 +/- 6.3, p>0.24) with limits of agreement ranging from +242 to -200 kcal. Resting energy expenditure measured by BG was significantly less than WRC values (accurate prediction: 47.4%; %bias: 11.0 +/- 14.6, p=0.0001) with unacceptably wide limits of agreement. Harris-Benedict, Schofield and World Health Organization equations were the most accurate, estimating REE within 10% of measured REE, but none seem appropriate to predict the REE of individuals.
Conclusion: There was a poor agreement between the REE measured by WRC compared to BG or predictive equations. The BG assessed REE accurately in 47.4% of the subjects on an individual level.