Abstract
ObjectivesTo compare the repeatability of Cone Location and Magnitude Index expanded (CLMI.X) parameters of 2 mm diameter zone of greatest corneal curvature (Cspot-Axi) and 1 mm diameter zone of thinnest pachmymetry (Spot-Pach) with the maximum single point keratometry (Kmax), 3 mm Zonal Kmax (Z-Kmax3), and thinnest single point pachymetry (TP) in keratoconus (KC).MethodsIn this Comparative repeatability study, data from 36 eyes of 36 normal individuals and 72 eyes of 72 KC patients (28 eyes with <= 50.0D and 44 eyes with >50.0D Z-Kmax3) were analyzed. For each enrolled eye, imaging was done 6 times (3 consecutive acquisitions with a half hour break). For each parameter, the within-subject standard deviation (Sw) was calculated from the data of the six exams.ResultsIn the normal group, Cspot-Axi-Sw was different from Kmax-Sw (p = 0.0004). Also, Z-Kmax3-Sw was different from Kmax-Sw (p = 0.0297). The difference between Cspot-Axi-Sw and Z-Kmax3-Sw was statistically significant (p = 0.0482). In the KC group, Sw were significantly different between Cspot-Axi and Kmax (p < 0.0001), and between Z-Kmax3 and Kmax (p < 0.0001). In the Z-Kmax3 <= 50.0D subgroup, Sw were different between Cspot-Axi and Kmax (p = 0.0002). In the Z-Kmax3 > 50.0D subgroup, Sw were different between Cspot-Axi and Kmax (p < 0.0001), and between Z-Kmax3 and Kmax (p < 0.0001). Sw differences between Spot-Pach and TP were not significant in any of the study groups (all P > 0.05).ConclusionsIn the diagnosis and follow up of KC, the zonal averages of Cspot-Axi and Z-Kmax3 are more reliable than the single point Kmax. The repeatability of Spot-Pach is similar to TP and both variables should suit the purpose equally.